7. Maintenance
7-4 Performing calibration
123 / 168
7-4-2 Performing zero calibration
Gas concentrations cannot be accurately measured unless zero calibration is performed correctly.
Be sure to carry out zero calibration to ensure accurate measurement.
Prepare the zero calibration gas, then connect to the product before starting zero calibration.
For information on zero calibration preparation, refer to ‘7-4-1 Calibration preparation’.
WARNING
Before performing zero calibration for the surrounding atmosphere, confirm that the surrounding
atmosphere contains clean air. Performing zero calibration in the presence of miscellaneous gases will
prevent correct calibration and create safety hazards in the event of an actual gas leak.
CAUTION
Perform zero calibration after supplying the zero calibration gas and waiting for the reading to stabilize.
Summary of Contents for GD-3EC
Page 169: ...11 Appendix Revision History 169 168...
Page 170: ...11 Appendix Revision History 170 168...
Page 171: ...11 Appendix Revision History 171 168...
Page 172: ...11 Appendix Revision History 172 168...
Page 173: ...11 Appendix Revision History 173 168...
Page 174: ...11 Appendix Revision History 174 168...
Page 175: ...11 Appendix Revision History 175 168...