10. Product Specifications
10-3 Detection target gases for individual detection principles
154 / 168
10-3 Detection target gases for individual detection principles
10-3-1 Non-dispersive infrared absorption type (Infrared type)
Sensor model
Gas name
Gas
symbol
F.S.
1 digit
First
Alarm
setpoint
Second
Alarm
setpoint
Calibration
gas
Operating
temperature
range
Operating
humidity
range
Years of
warranty
IRF-1301*
Methane
CH
4
100 %LEL
0.5 %LEL
25 %LEL
50 %LEL
CH
4
−40 °C -
70 °C
95 %RH or
less
(no conden-
sation)
3 years
IRF-1303*
Isobutane
i-C
4
H
10
100 %LEL
0.5 %LEL
25 %LEL
50 %LEL
i-C
4
H
10
IRF-1315*
Propane
C
3
H
8
100 %LEL
0.5 %LEL
25 %LEL
50 %LEL
C
3
H
8
i-C
4
H
10
IRF-1435
Carbon dioxide
CO
2
2,000 ppm
1 ppm
1,000 ppm
-
CO
2
IRF-1436
Carbon dioxide
CO
2
5,000 ppm
10 ppm
1,000 ppm
-
CO
2
IRF-1433
Carbon dioxide
CO
2
10,000 ppm
10 ppm
1,000 ppm
-
CO
2
IRF-1437
Carbon dioxide
CO
2
2 vol%
0.005 vol%
1 vol%
-
CO
2
IRF-1438
Carbon dioxide
CO
2
5 vol%
0.010 vol%
1 vol%
-
CO
2
IRF-1439
Carbon dioxide
CO
2
10 vol%
0.01 vol%
1 vol%
-
CO
2
IRF-1334*
Methane
CH
4
100 vol%
0.5 vol%
25 vol%
50 vol%
CH
4
IRF-1316*
Ethylene
C
2
H
4
100 %LEL
0.5 %LEL
25 %LEL
50 %LEL
C
2
H
4
CH
4
IRF-1340*
Isobutylene
i-C
4
H
8
100 %LEL
0.5 %LEL
25 %LEL
50 %LEL
i-C
4
H
8
i-C
4
H
10
IRF-1308*
n-hexane
n-C
6
H
14
100 %LEL
0.5 %LEL
25 %LEL
50 %LEL
n-C
6
H
14
i-C
4
H
10
IRF-1332*
Butadiene
C
4
H
6
100 %LEL
0.5 %LEL
25 %LEL
50 %LEL
C
4
H
6
CH
4
*Interference occurs due to other hydrocarbon gases.
NOTE
Install in a location with stable conditions within the specified operating temperature range and free of
sudden fluctuations.
When calibrating, allow the product to warm up for at least 10 minutes before performing zero calibration
and span adjustment.
Make sure temperature, humidity, and pressure conditions in the calibration environment are the same as
for the measurement environment. (The reading may fluctuate due to temperature, humidity, and
pressure characteristics and in some cases may exceed the zero suppression value due to sensor
characteristics.)
For ethylene or butadiene to which the product has low gas sensitivity, set the alarm setpoint to 25 %LEL
or higher.
Due to infrared absorption characteristics, readings may differ even for the same target gas concentration
if the base gas composition differs significantly between calibration and measurement—for example, if an
air base is used for calibration and a CO
2
base is used for measurement.
Perform maintenance if a warning appears while the life assessment or self-diagnostic function is active.
Regardless of the detection target gas, the warranty period for IRF sensors is 3 years.
Summary of Contents for GD-3EC
Page 169: ...11 Appendix Revision History 169 168...
Page 170: ...11 Appendix Revision History 170 168...
Page 171: ...11 Appendix Revision History 171 168...
Page 172: ...11 Appendix Revision History 172 168...
Page 173: ...11 Appendix Revision History 173 168...
Page 174: ...11 Appendix Revision History 174 168...
Page 175: ...11 Appendix Revision History 175 168...