RF627, RF627Smart [Revision 2.1.2] 20.09.2021
Laser Scanners
37
Parameter
Factory value
Description
Divider
is "2", it applies for every second signal at the input, etc.
Setting the
Divider
parameter allows, for example, to match the
frequency of the input signals and the permissible frequency of the
scanner.
Delay, us
OFF
Delay from the start of the synchronization signal to the
synchronization event (the start of the measurement cycle).
20.4.1. Setting the inputs
The
Inputs
section contains the parameters of the inputs.
Parameters of Inputs #1 and #2:
Parameter
Factory value
Description
Enable
OFF
Enable / disable the input.
Mode
Rise or Fall
Input signal processing mode:
Rise or fall – Synchronization by rise or fall.
Rise – Synchronization by rise.
Fall – Synchronization by fall.
High level — Synchronization by high level.
Low level — Synchronization by low level.
Input #3 of the scanner is designed to connect the reset signal of the measurement
counters and the encoder.
Parameters:
Parameter
Factory value
Description
Enable
OFF
Enable / disable the input.
Mode
Rise
Reset signal processing mode:
Rise — Reset by front.
Fall — Reset by fall.
20.4.2. Setting the encoder counter
The
Counter
section contains the settings for the encoder counter.