RF627, RF627Smart [Revision 2.1.2] 20.09.2021
Laser Scanners
31
The
Max Intensity
value determines the selection of the profile point based on the maximum brightness of
the image in the CMOS sensor column. The brightness of the re-reflected signal may be greater than the
brightness of the original signal. The scanner incorrectly selects the profile, placing it both on the initial laser
line and on the re-reflexion.
The
First
value determines the selection of the first peak in the CMOS sensor column.
The scanner selects a profile by a re-reflected signal.
The
Last
value determines the selection of the last peak in the CMOS sensor column.
The scanner selects a profile by a real signal.