Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
180
Operating Manual 1176.7561.02 ─ 06
Parameters:
<Level>
<numeric value>
Range:
70.711 nV to 7.071 V
*RST:
1 V
Example:
INP:DIQ:RANG 1V
Mode:
A, IQ, NF, TDS, VSA, CDMA, EVDO, WCDMA, ADEMOD, GSM,
OFDM, OFDMA/WiBro, WLAN
Manual operation:
See
INPut:DIQ:RANGe[:UPPer]:UNIT
<Unit>
Defines the unit of the full scale level (see
on page 78). The availability of
units depends on the measurement application you are using.
This command is only available if the optional R&S Digital I/Q Interface (option
R&S
FSV-B17) is installed.
For details see the R&S Digital I/Q Interface (R&S
FSV-B17) description of the base
unit.
Parameters:
<Level>
V | dBm | dBpW | W | dBmV | dBuV | dBuA | A
*RST:
Volt
Example:
INP:DIQ:RANG:UNIT A
Mode:
IQ, VSA, EVDO, CDMA, WCDMA, GSM, ADEMOD, TDS
Manual operation:
See
INPut:DIQ:SRATe
<SampleRate>
This command specifies or queries the sample rate of the input signal from the R&S
Digital I/Q Interface (see
Note:
the final user sample rate of the R&S
FSVA/FSV may differ and is defined using
SENSe:ADEM:SRATe
(see
This command is only available if the optional R&S Digital I/Q Interface (option
R&S
FSV-B17) is installed.
For details see the R&S Digital I/Q Interface (R&S
FSV-B17) description of the base
unit.
Parameters:
<SampleRate>
Range:
1 Hz to 10 GHz
*RST:
32 MHz
Example:
INP:DIQ:SRAT 200 MHz
Mode:
A, IQ, NF, TDS, VSA, CDMA, EVDO, WCDMA, ADEMOD, GSM,
OFDM, OFDMA/WiBro, WLAN
Remote Commands of the Analog Demodulation (R&S