Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
143
Operating Manual 1176.7561.02 ─ 06
Parameters:
<Threshold>
numeric value
The absolute threshold level
Range:
-150 dBm to 30 dBm
*RST:
-40 dBm
Example:
DEM:SQU:LEV -80
If the signal drops below -80 dBm, the demodulated data is set
to 0.
Usage:
SCPI confirmed
Mode:
ADEMOD, SFM
Manual operation:
See
[SENSe:]ADEMod:SRATe?
This command returns the sample rate set up for the current analog demodulation
measurement.
Example:
ADEM:SRAT?
Returns the current sample rate.
Usage:
Query only
Mode:
ADEMOD
[SENSe:]ADEMod<n>:ZOOM[:STATe]
<State>
The command enables or disables the zoom function for the analog-demodulated mea-
surement data in the specified window. Depending on the selected measurement time
and the demodulation bandwidth, the number of recorded test points may be greater
than that shown on the display.
If the zoom function is enabled, the default number of sweep points in "Spectrum"
mode of the result memory are displayed from the specified start time with
If the zoom function is disabled, data reduction is used to adapt the test points to the
number of points available on the display.
Suffix:
<n>
.
1...4
window
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
ADEM:ZOOM ON
Switches on the zoom function
Mode:
ADEMOD
Remote Commands of the Analog Demodulation (R&S