Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
105
Operating Manual 1176.7561.02 ─ 06
Example:
CALC:MARK3 ON
Switches on marker 3 or switches to marker mode.
Manual operation:
See
"Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
See
CALCulate<n>:MARKer<m>:TRACe
<Trace>
This command selects the trace a marker is positioned on.
The corresponding trace must have a trace mode other than "Blank".
If necessary, the corresponding marker is switched on prior to the assignment.
Suffix:
<n>
.
Selects the measurement window.
<m>
depends on mode
Selects the marker.
Parameters:
<Trace>
1 ... 6
Trace number the marker is positioned on.
Example:
CALC:MARK3:TRAC 2
Assigns marker 3 to trace 2.
Manual operation:
See
See
CALCulate<n>:MARKer<m>:X
<Position>
This command positions a marker on a particular coordinate on the x-axis.
If marker 2, 3 or 4 is selected and used as delta marker, it is switched to marker mode.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Parameters:
<Position>
Numeric value that defines the marker position on the x-axis.
The unit is either Hz (frequency domain) or s (time domain) or
dB (statistics).
Range:
The range depends on the current x-axis range.
Example:
CALC:MARK2:X 1.7MHz
Positions marker 2 to frequency 1.7 MHz.
Manual operation:
See
"Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
Remote Commands of the Analog Demodulation (R&S
FSV–K7)