Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
139
Operating Manual 1176.7561.02 ─ 06
<trigger slope>
POSitive | NEGative
Used slope of the trigger signal.
The value indicated here will be ignored for <trigger source> =
IMMediate.
*RST:
POSitive
<offset samples>
Number of samples to be used as an offset to the trigger signal.
For details refer to
Chapter 2.1.3, "Sample Rate, Measurement
The value indicated here is ignored for <trigger source> =
"IMMediate".
*RST:
0
<# of meas>
Number of repetitions of the measurement to be executed. The
value indicated here is especially necessary for the average/
maxhold/minhold function.
Range:
0 to 32767
*RST:
0
Example:
ADEM:SET 8MHz,32000,EXT,POS,-500,30
Performs a measurement at:
sample rate = 8 MHz
record length = 32000
trigger source = EXTernal
trigger slope = POSitive
offset samples = -500 (500 samples before trigger occurred)
# of meas = 30
Mode:
ADEMOD
[SENSe:]ADEMod:SPECtrum[:TYPE]
<Type>
This command selects the result types to be created in parallel by the RF spectrum
measurement with active analog demodulation.
Parameters:
<Type>
*RST:
OFF,OFF,OFF
<result type 1|2|3|4|5|6>: WRITe, AVERage, MAXHold, MINHold,
VIEW, OFF; for details see
Chapter 2.3.6.1, "Trace Mode Result
Example:
ADEM:SPEC AVER,MAXH,MINH
Creates average, max hold and min hold values at a time
ADEM:SPEC WRIT,OFF,OFF
Only creates the current measurement values
ADEM:SPEC OFF,OFF,OFF
Switches analog demodulator off
Mode:
ADEMOD
Remote Commands of the Analog Demodulation (R&S
FSV–K7)