FM Stereo Option R&S
FSV–K7S
R&S
®
FSV-K7/K7S
209
Operating Manual 1176.7561.02 ─ 06
Phase Unit (Rad/Deg) ← Unit
Sets the phase unit to rad or deg for displaying PM signals.
Remote command:
THD Unit (% / DB) ← Unit
Sets the unit to percent or DB for THD measurements.
Remote command:
Abs. Dev Unit (kHz/dBm) ← Unit
Sets the unit for absolute deviation to kHz or dBm. This softkey is only available with
the FM Stereo option K7S.
Remote command:
Rel. Dev Unit (dB / %) ← Unit
Sets the unit for relative deviation to dB or percent. This softkey is only available with
the FM Stereo option K7S.
Remote command:
Preamp On/Off
Switches the preamplifier on and off.
If option R&S
FSV-B22 is installed, the preamplifier is only active below 7 GHz.
If option R&S
FSV-B24 is installed, the preamplifier is active for all frequencies.
This function is not available for input from the R&S Digital I/Q Interface (option
R&S
FSV-B17).
Remote command:
RF Atten Manual/Mech Att Manual
Opens an edit dialog box to enter the attenuation, irrespective of the reference level. If
electronic attenuation is activated (option R&S
FSV-B25 only; "El Atten Mode Auto"
softkey), this setting defines the mechanical attenuation.
The mechanical attenuation can be set in 10 dB steps.
The RF attenuation can be set in 5 dB steps (R&S
FSV with option R&S
FSV-B25 or
R&S
FSVA: 1 dB steps). The range is specified in the data sheet. If the current refer-
ence level cannot be set for the set RF attenuation, the reference level is adjusted
accordingly.
This function is not available for input from the R&S Digital I/Q Interface (option
R&S
FSV-B17).
The RF attenuation defines the level at the input mixer according to the formula:
level
mixer
= level
input
– RF attenuation
Instrument Functions FM Stereo (R&S
FSV–K7S)