PROBe:SETup:MODE
<Mode>
This command defines which action is taken when the probe's micro button is pressed.
Parameters:
<Mode>
RSINgle
A single sweep is performed.
NOACtion
No action is taken.
*RST:
OFF
Example:
PROB:SET:STAT ON
Manual operation:
See
PROBe:SETup:NAME?
This command returns the name of the connected probe.
Example:
PROB:SET:NAME?
Usage:
Query only
PROBe:SETup:STATe?
This command queries whether a probe is connected to the instrument's [RF Input] and
USB connectors and was recognized by the R&S
FSVA/FSV.
Example:
PROB:SET:STAT ON
Usage:
Query only
PROBe[:STATe]
<State>
This command activates a connected probe. Use this command to switch off the probe
and measure the digital input without considering the transducer factor of the probe.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
PROB:STAT ON
Manual operation:
See
2.3.6
SENSe Subsystem (Analog Demodulation, R&S
FSV–K7)
The SENSe subsystem is organized in several subsystems. The commands of these
subsystems directly control device-specific settings, they do not refer to the signal
characteristics of the measurement signal.
Remote Commands of the Analog Demodulation (R&S