Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
137
Operating Manual 1176.7561.02 ─ 06
Example:
ADEM:SET 8MHz,32000,EXT,POS,-500,30
, see
Sets demodulator
ADEM:PM AVER,MAXH,MINH
, see
Selects the PM results to be measured
ADEM:AM:REL WRIT,OFF,OFF
, see
Selects the AM results to be measured
ADEM:PM:AFSP WRIT,OFF,OFF
, see
Selects the AF spectrum results of the demodulated PM signal
to be measured
ADEM ON
, see
Switches the demodulator on
INIT; WAI
Starts the measurement and waits for termination
FORM ASC
, see
Selects output format
ADEM:PM:RES? AVER
, see
Reads PM average result data
ADEM:PM:RES? MAXH
, see
Reads PM maxhold result data
ADEM:PM:RES? MINH
, see
Reads PM minhold result data
ADEM:AM:RES? WRIT
, see
Reads current AM result data
ADEM:PM:AFSP:RES? WRIT
Reads current AF spectrum result data of demodulated PM sig-
nal
Mode:
ADEMOD
[SENSe:]ADEMod:PM:RPOint[:X]
<Time>
This command determines the position where the phase of the PM-demodulated signal
is set to 0 rad. The maximum possible value depends on the measurement time
selected in the instrument; this value is output in response to the query
ADEM:PM:RPO:X? MAX
.
Parameters:
<Time>
0 s to measurement time
*RST:
0 s
Example:
ADEM:PM:RPO 500us
Sets the position where the phase to 0 rad setting to 500 µs.
Remote Commands of the Analog Demodulation (R&S
FSV–K7)