Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
184
Operating Manual 1176.7561.02 ─ 06
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
OUTP:DIQ ON
Mode:
ADEMOD, IQ, VSA
OUTPut:IF[:SOURce]
<Source>
This command selects the source of the IF output.
Parameters:
<Source>
IF
Outputs the intermediate frequency.
VIDeo
Outputs the video signal (200 mV).
*RST:
IF
Example:
OUTP:IF VID
Selects the video signal for the IF output connector.
Manual operation:
See
OUTPut:TRIGger
<PortLevel>
This command selects level of the Trigger Out port. Thus, you can trigger an additional
device via the external trigger port, for example.
Parameters:
<PortLevel>
LOW | HIGH
*RST:
LOW
Example:
OUTP:TRIG HIGH
Manual operation:
See
2.3.10.5
Other Commands
FORMat[:DATA]
<Format>
This command selects the data format for the data transmitted from the
R&S
FSVA/FSV to the controlling computer. It is used for the transmission of trace
data. The data format of trace data received by the instrument is automatically recog-
nized, regardless of the format which is programmed.
(See also
Parameters:
<Format>
ASCii
ASCII data are transmitted in plain text, separated by commas.
Remote Commands of the Analog Demodulation (R&S
FSV–K7)