Analog Demodulation Option R&S
FSV–K7
R&S
®
FSV-K7/K7S
175
Operating Manual 1176.7561.02 ─ 06
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Parameters:
<TraceNumber>
1 ... 6
Trace number the marker is positioned on.
Example:
CALC:DELT3:TRAC 2
Assigns delta marker 3 to trace 2.
Manual operation:
See
CALCulate<n>:DELTamarker<m>:X
<Position>
This command positions a delta marker on a particular coordinate on the x-axis.
The position is an absolute value.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Parameters:
<Position>
0 to maximum frequency or sweep time
Example:
CALC:DELT:X?
Outputs the absolute frequency/time of delta marker 1.
Manual operation:
See
"Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
CALCulate<n>:DELTamarker<m>:X:RELative?
This command queries the x-value of the selected delta marker relative to marker 1 or
to the reference position (for
CALC:DELT:FUNC:FIX:STAT ON
). The command acti-
vates the corresponding delta marker, if necessary.
Suffix:
<n>
.
Selects the measurement window.
<m>
Selects the marker.
Example:
CALC:DELT3:X:REL?
Outputs the frequency of delta marker 3 relative to marker 1 or
relative to the reference position.
Usage:
Query only
Manual operation:
See
"Marker 1 / Marker 2 / Marker 3 / … Marker 16,/ Marker
Remote Commands of the Analog Demodulation (R&S