The SENSe subsystem controls the essential parameters of the analyzer. In accord-
ance with the SCPI standard, the keyword "SENSe" is optional for this reason, which
means that it is not necessary to include the SENSe node in command sequences.
The following subsystems are included:
2.3.6.1
2.3.6.2
Formats for Returned Values: ASCII Format and Binary Format................................ 118
2.3.6.3
SENSe:ADEMod Subsystem (Analog Demodulation, Analog Demodulation)............ 119
2.3.6.4
2.3.6.5
SENSe:BANDwidth Subsystem (Analog Demodulation, R&S
FSV–K7).....................147
2.3.6.6
SENSe:FILTer Subsystem (Analog Demodulation, Analog Demodulation)................ 148
2.3.6.7
SENSe:FREQuency Subsystem (Analog Demodulation, R&S
FSV–K7)................... 153
2.3.6.8
SENSe:SWEep Subsystem (Analog Demodulation, R&S
FSV–K7)...........................155
2.3.6.9
2.3.6.1
Trace Mode Result Types
The following result types can be set:
WRITe
The current trace results will be obtained
AVERage
The trace results will be averaged over the given # of measurements
MAXHold
The maximum trace result values will be obtained over the given # of measurements
MINHold
The minimum trace result values will be obtained over the given # of measurements
VIEW
The trace results are frozen and displayed, i.e. they are not calculated for subse-
quent measurements. Traces in this mode cannot be queried.
OFF
The result type will not be used.
It is not possible to query trace data when result type VIEW is selected.
Each value besides OFF can only be assigned to one result type at a time.
If all result types are set to OFF, the AM, FM, or PM demodulator will be deactivated.
2.3.6.2
Formats for Returned Values: ASCII Format and Binary Format
●
ASCII Format (FORMat ASCII):
The command reads out a list of comma separated values (CSV) of the measured
values in floating point format.
●
Binary Format (FORMat REAL,32):
The command reads out binary data (Definite Length Block Data according to IEEE
488.2), each measurement value being formatted in 32 Bit IEEE 754 Floating-
Point-Format. The schematics of the result string will be as follows:
#41024<value1><value2>…<value n> with
Remote Commands of the Analog Demodulation (R&S
FSV–K7)