2461Type C User Manual
Racal Instruments
1998
Programming The Type C Function Card 4-51
SENSE
SUBSYSTEM
The commands in the SENSe subsystem perform the required
configuration of the 2461 TYPE C instrument (measurement
function and other settings). The reconfiguration of the
measurement function invalidates the previously acquired data.
If the SENSe:FUNCtion command is sent while the 2461 TYPE
C is initiated, then measurement in progress is aborted before
the reconfiguration takes place SENSe command summary
COMMAND
DESCRIPTION
[:SENSE[1 | 2 | 3]]
:FUNCTION
[SENSe]
[:PERiod]:APERture
[:FREQuency]:APERture
[:FREQuency:RATio]:APERture
[:TINTerval]:APERture
[:RTIMe]:APERture
[:RISE:TIME]:APERture
[:FTIMe]:APERture
[:FALL:TIME]:APERture
[:PWIDth]:APERture
[:NWIDth]:APERture
[:PHASe]:APERture
[:TOTalize:APERture
[:CHECk]:APERture
:AVERAGE[:STATE]
:TINTerval
:DELay
:TIME
[:STATe]
:TOTalize
:GATe
[:STATe]
:POLarity
:INTermediate
[:STATe]
:MANual
:RESolution
Selects measurement function:
″
PERiod”,
″
FREQuency
″
,
″
FREQuency:RATio
″
,
″
TINTerval
″
,
″
RTIMe
″
,
″
RISE:TIME
″
,
″
FTIMe
″
,
″
FALL:TIME
″
,
″
PWIDth
″
,
″
NWIDth
″
,
″
PHASe
″
,
″
CHECk
″
,
″
TOTalize
″
,
″
VOLTage:MAXimum
″
, “VOLTage:MINium
″
,
″
VOLTage:MIDDle
″
Selects the measurement aperture (gate time). The function
parameter is optional because the aperture value applies to
all functions.
Enables or disables average.
Selects time delay.
Enables or disables time interval delay.
Selects TOTalize 1 by 2, 2 by 1 or Manual Totalize.
Selects slope of channel.
Enables or disables intermediate results of totalize function
Starts and stops Manual Totalize
Specifies number of digits of resolution.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com