User Manual 2461 Type C
A-2
INPUT 1 in COMMON mode:
1M
Ω
nominal shunted by
≤
55pF
or
50
Ω
(DC coupled)
Sensitivity
×
1 attenuation
Sinewave:
10Hz
≤
f
≤
100MHz
25mV RMS
100MHz < f
≤
160MHz
50mV RMS
160MHz < f
≤
200MHz
75mV RMS
where: f is the input frequency in Hz
Pulse:
75mV for pulse widths down to 5ns
Variable Sensitivity (Hysteresis)
Two selectable settings of the hysteresis level allow the sensitivity to be adjusted. The
above Sensitivity figures are with the hysteresis in its most sensitive position, but the
sensitivity will be reduced by a factor of 2 nominal upon selecting the other setting.
Dynamic Range
x1 attenuation:
10Hz
≤
f
≤
50MHz
5Vpk-pk min.
50MHz < f
≤
160MHz
2. 5V pk-pk min.
160MHz < f
≤
200MHz
1.5Vpk-pk min.
x10 attenuation:
10Hz
≤
f
≤
50MHz
50Vpk-pk min.
50MHz < f
≤
160MHz
25Vpk-pk min.
160MHz < f
≤
200MHz
15Vpk-pk min.
where: f is the input frequency in Hz)
Signal Operating Range
x1 attenuation:
-5.1VDC to +5.1VDC