2461Type C User Manual
Racal Instruments
1998
Performance Verification and Adjustments 6-3
Frequency Signal
level
Resolution
Command
Syntax
Tolerance
10Hz
25mV rms
SENS:RES 8
± 1Hz
5kHz
25mV rms
SENS:RES 8
± 0.1Hz
10kHz
25mV rms
SENS:RES 8
± 0.1Hz
Set the UCT measurement resolution to the values given in the
following table and apply the Frequencies given to the UCT
input 1. Check that the UCT gives a stable reading within the
tolerances given, at less than the signal levels given.
Table 6-2, Input 1 Sensitivity high Frequency Performance Limits
Frequency Signal
level
Resolution
Command
Syntax
Tolerance
200MHz
50mV rms
SENS:RES 9
± 1Hz
160MHz
50mV rms
SENS:RES 9
± 1Hz
100kHz
50mV rms
SENS:RES 9
± 1Hz
10MHz
25mV rms
SENS:RES 8
± 1Hz
100kHz
25mV rms
SENS:RES 8
± 0.1Hz
Disconnect the test equipment.