2461Type C User Manual
Performance Verification and Adjustments 6-8
Racal Instruments
1998
Time Interval 1:2
Test
Set the UCT to its default state. Select Time Interval 1:2 as the
measurement mode, 50
Ω
impedance for input 1, Common
mode for inputs 1 and 2 and External frequency standard.
Equipment: Signal Generator
Configure the UCT as follows:
*RST
returns unit to default state.
INP1:IMP
50
selects
50
Ω
impedance
IMP1:ROUT COMM
commons Channels1 and 2
selects external frequency
standard
CONF1:TINT
selects time Interval 1-2
If the UCT is fitted with a TCXO or OCXO oscillator option,
connect the UCT internal reference output to the external signal
source 10MHz Input. Configure the external signal source to
accept the 10MHz reference from the UCT.
Connect the Function Generator to the UCT as shown in Figure
6-3.
Figure 6-3, Time Interval 1-2 Test Configuration
Set the UCT input trigger slopes as shown in the following table
and check that the UCT measurement results are as given in
Table 6-5.
GATE
IN
1
VT1
VT2
ARM
T2
IN
STD
OUT
IN
2
OSC
UCT
IN 3
SIGNAL GENERATOR
RF OUT
10MHz OUT
INPUT 1
EXTERNAL STANDARD INPUT