2461Type C User Manual
Programming The Type C Function Card 4-54
Racal Instruments
1998
Parameter:
MIN = 1E-6 s
MAX = 10s
DEFault = 1s
1E-6 < value < 10
Initialization:
*RST 0.1s
Example:
″
SENSe:APERture 0.6
″
The aperture value is rounded to 1s
Interrogative form:
[:SENSe]:APERture? ==> <aperture time>
Returns the aperture time selected.
*************************************************************
[SENSe]:AVERage[:STATe] <ON | OFF>
Description:
Enables or disables the averaging of measurement results.
When the average function is enabled, the 2461 TYPE C takes
100 measurements and derives the average value. Note that
the resolution increases by one digit when this function is
enabled.
NOTE:
This function is only available with frequency (on
channel 1 and channel 2), period, ratio and time interval
measurements.
Initialization:
*RST OFF
Example:
″
SENSe:AVERage ON
″
Enables averaging.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com