![background image](http://html2.mh-extra.com/html/racal-instruments/2461/2461_user-manual_3134747102.webp)
2461Type C User Manual
Racal Instruments
1998
Programming The Type C Function Card 4-59
FETCh? ---> 155
Returns an intermediate
count
SENSe2:TOTalize:MANual OFF
Stops the counting
FETCh? ->>235
Returns the final count
SENSe2:TOTalize:MANual ON Re-starts the counting
FETCh? ---> 275
Returns an intermediate
count
SENSe2:TOTalize:MANual OFF
Stops the counting
FETCh? ->>395
Returns the final count
SENSe:TOTalize:GATE OFF Resets the count
Interrogative form:
SENSe1:TOTalize:MANual? ==> <ON | OFF>
Returns the current state of the Manual Totalize
function.
*******************************************************
[SENSe]:RESolution <value | MINimum | maximum |
DEFault>
Description:
Selects the number of digits of resolution. This is in the range
of 3 to 10.
In frequency, period, ratio and check modes, the gate time is set
by the resolution selected. The resolution of phase and totalize
measurements is determined by the input signal.
The resolution setting may be changed indirectly by the
CONFigure:ARRay and MEASure:ARRay commands and by
the SENSe:APERture command.
Initialization:
*RST 9
Parameters:
3
≤
value
≤
10
MIN = 3
MAX = 10
DEF = 8
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com