Theory of Operation 3-15
Capabilities
Arming of the START and STOP circuits can be either internally
or externally derived. External arming is possible on all functions
except phase. The following arming modes are available:
Mode Start
Stop
1
INT ARM
INT ARM
2
EXT ARM
↑
INT ARM
3
EXT ARM
↓
INT ARM
4 INT
ARM
EXT ARM
↑
5 INT
ARM
EXT ARM
↓
6
EXT ARM
↑
EXT
ARM
↑
7
EXT ARM
↑
EXT
ARM
↓
8
EXT ARM
↓
EXT
ARM
↑
9
EXT ARM
↓
EXT
ARM
↓
The EXT ARM input can be from any of the VXI TTLO-7 lines,
or from the front panel ARM input.
For modes that use INT ARM, the STOP arming hold-off time is
adjustable for time interval measurements.
Correct use of the arming facilities can be used, for example, to
prevent premature triggering of the START and/or STOP circuits
due to external noise (e.g. contact bounce).
EXTERNAL FREQUENCY REFERENCE INPUT
The 2461 Type C can make measurements using as its
frequency reference an external signal connected to the front
panel port marked STD IN.
INTERNAL FREQUENCY REFERENCE OUTPUT
The frequency reference in use is available at the front panel