2461Type C User Manual
Performance Verification and Adjustments 6-10
Racal Instruments
1998
Totalize 1 by 2 Test
Set the UCT to its default state. Select Totalize 1 by 2 as the
measurement mode, 50
Ω
input 1 impedance, and external
frequency standard.
Equipment: Signal Generator
Configure UCT as follows:
*RST
return unit to default state
INP1:IMP
50
selects
50
Ω
impedance for
Input 1
SOUR:ROSC:SOUR EXT
selects external frequency
standard
If the UCT is fitted with a TCXO or OCXO oscillator option,
connect the UCT internal reference output to the external signal
source 10MHz Input. Configure the external signal source to
accept the 10MHz reference from the UCT.
Connect the UCT to the signal generator as shown in Figure 6-
6. Set the signal generator to 10MHz, 0dBm. Check that the
UCT measurement result is 5000 +/- 1.
Figure 6-4, Totalize 1 by 2 Test Configuration
Command Syntax for making a Totalize measurement:
CONF1:TOT
configures UCT for Totalize
measurement
MEAS1:TOT?
takes a totalize measurement
GATE
IN
1
VT1
VT2
ARM
T2
IN
STD
OUT
IN
2
OSC
UCT
IN 3
SIGNAL GENERATOR
RF OUT
10MHz OUT
INPUT 1
EXTERNAL STANDARD INPUT
INPUT 2
1KHz AF OUT