2461Type C User Manual
Racal Instruments
1998
Programming The Type C Function Card 4-75
ignored”).
Initialization:
*RST IMMediate
Example:
″
ARMing:STOP:SOURce TTLTrg2
″
Selects a TTL trigger line 2 for the stop arming.
Interrogative form:
ARMing: (SEQuence2 | STOP)[:LAYer]:SOURce? ==>
<CTR_EXTARM
|
TTLTrg<0..7> |
IMMediate>
Returns the source of the stop arming signal.
2461 TYPE C-
Specific Error
Messages
Error Description [Description/Explanation/Examples]
Number
-311 Memory-related error
[An error has occurred in one of the memory areas of the
module. The problem may arise when programming the Non-
volatile memory or when transferring samples in DMA mode.]
-340 Calibration failed
[An error has occurred during the calibration process]
-350 Queue Overflow
[A specific code has been entered in the Queue instead of the
code, which caused the error. This code indicates that the
Queue is full and that the error that occurred has not been
recorded.]
Examples Of SCPI Commands Used To Manipulate The
Instrument
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com