User Manual 2461 Type C
A-4
Low Pass Filter
Independently selectable in Channel 1 and Channel 2.
3dB bandwidth:
50kHz nominal
Attenuation rate:
20dB per decade nominal
Input Connection
BNC female
Crosstalk
> 36dB separation between channels 1 & 2
(measured at 100MHz in a 50
Ω
system,
×
1 attenuation).
Trigger Slope
Independently selectable positive or negative.
Manual Trigger
(1)
Trigger Level Range
-5.1V to +5.1VDC
(2)
Trigger Level Resolution
x1 attenuation
2.5mV nominal
x10 attenuation
25mV nominal
(3)
Trigger Level Accuracy
(Relative to the trigger point as queried with INP[1|2]:COMP:LEV?)
×
1
±
1% reading
±
30mV
×
10
±
1% reading
±
300mV
(4)
Trigger Level Output
Independent buffered outputs representing the trigger level.
Range:
±
5.1VDC irrespective of attenuator setting. For X10 the
measured output voltage should be scaled by a factor of 10.
Resolution:
2.5mV nominal.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com