Programming The Type C Function Card 4-23
CONFigure?
DESCRIPTION:
This query returns the current function, channel and required
number of measurements. The function strings returned are:
“FREQ”
“FREQ:RAT”
“VOLT:MAX”
“VOLT:MIN”
“VOLT:MIDD”
“PWID”
“NWID”
“PER”
“RTIM”
“FTIM”
“TINT”
“TOT”
“PHAS”
“CHEC”
EXAMPLE:
″
CONFigure?
″
This query would return the string “
FREQ1,5
” if the
function selected is FREQuency, the current channel is
1 and an ARRay-type measurement (5 samples) was
specified.
Initialization:
*RST
FREQ1,1