Programming The Type C Function Card 4-41
*RST AC
INTERROGATIVE FORM
INPut[1 | 2]:COUPling? ==> <AC | DC>
Returns the current coupling setting for the specified
channel.
*******************************************************
INPut[1 | 2]:IMPedance <value | MINimum | MAXimum |
DEFault>
DESCRIPTION:
Selects the input terminating impedance for the specified
channel. If <value> is less than 60, the impedance is set to
50
Ω
. If <value> is greater than 60, the impedance is set to
1M
Ω
.
PARAMETERS:
0
≤
value
≤
10E6
Ω
MIN = 50
Ω
MAX = 1E6
Ω
DEFault = 1E6
Ω
INITIALIZATION:
*RST MAX
INTERROGATIVE FORM
INPut[1 | 2]:IMPedance? ==> <50 | 1000000>
Returns the current impedance setting for the specified
channel.
*******************************************************
INPut[1 | 2]:ROUTe <SEParate | COMMon | DEFault>
DESCRIPTION: