Appendix E
| Electromagnetic Compatibility Guidance
©2007-2015 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
263
Electromagnetic Compatibility Guidance
Electromagnetic Emissions
Table 47
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The LIFEPAK 15 monitor/defibrillator is intended for use in the electromagnetic environment
specified below. The customer or the user of the LIFEPAK 15 monitor/defibrillator should
assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic Environment - Guidance
RF emissions
CISPR 11
Group 1
The LIFEPAK 15 monitor/defibrillator uses RF energy only for
its internal function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The LIFEPAK 15 monitor/defibrillator is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage power
supply network that supplies buildings used for domestic
purposes.
Harmonic
emissions
IEC 61000-3-2
Not applicable
Voltage
fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......