Noninvasive Pacing
146
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION
No muscle response to
pacing
Patient’s heart rate may be
greater than noninvasive pacer
ppm
•
No action needed.
The Test Load is connected to
therapy cable
•
Remove the Test Load and
connect therapy electrodes to
cable.
Patient muscle response is
variable and depends on patient
condition. Muscular response to
pacing is not a reliable indicator
of current delivered.
•
No action needed.
Capture does not occur with
pacing stimulus
Current (mA) set too low
•
Increase pacing current.
(Administer sedation or
analgesia as needed.)
CONNECT CABLE
or
PACING
STOPPED
message appears
Therapy cable damaged
•
Replace therapy cable and
perform daily checks per
Operator’s Checklist.
CONNECT ELECTRODES
message appears
Pacing cable or electrode
disconnected
•
Reconnect and set current.
Electrodes not adhering to skin
•
Prepare skin.
Therapy cable damaged
•
Replace therapy cable and
perform daily checks per
Operator’s Checklist.
Electrodes outdated
•
Replace electrodes and set
current.
PACING IN PROGRESS
message appears
CPR
pressed
•
Press
PACER
to stop pacing,
if appropriate, and then press
CPR
.
Pacing stops spontaneously
and
PACER FAULT
message
appears
Internal error detected
•
Cycle power and start pacing
again.
•
Obtain service by a qualified
service technician.
Intrinsic QRS complexes not
sensed when pacing
ECG size too low
•
Increase ECG size or select
another lead.
Intrinsic QRS complexes are
occurring during pacemaker's
refractory period
•
Adjust PPM.
Pacing starts spontaneously
Patient’s heart rate falls below
set pacing rate
•
Appropriate pacemaker
function; assess patient.
During standby pacing, ECG
lead disconnects and pacing
begins asynchronously
•
Reconnect ECG lead.
Set pacing rate (ppm) and
ECG paced rate do not
appear to match
Internal error detected
•
Print ECG and calculate the
pace rate.
Improper sensing
(for example, sensing on
T-waves)
QRS complex too small
•
Select another lead.
T-wave too large
•
Adjust ECG size.
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......