Automated External Defibrillation (AED)
122
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
The
ANALYZING NOW—STAND CLEAR
prompts occur. The SAS analyzes the
patient’s ECG in
approximately 6 to 9 seconds and advises
either
SHOCK ADVISED
or
NO SHOCK
ADVISED
.
7.
Continue to follow the screen messages and voice prompts provided by the AED.
Shock Advised
The following prompts occur when shock is advised:
If the AED detects a shockable rhythm,
the
SHOCK ADVISED
prompts occur.
Charging to the joule setting for Shock
#1 begins. A charging bar appears and a
ramping tone sounds.
When charging is complete, the available
energy is displayed.
The
STAND CLEAR, PUSH SHOCK
BUTTON!
( ) message occurs, followed
by a “Shock ready” tone.
Clear everyone away from touching the
patient, bed, or any equipment that is
connected to the patient.
Press (shock) to deliver energy to the
patient.
When the (shock) button is pressed, the
ENERGY DELIVERED
message occurs
indicating that the energy transfer was
completed.
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......