Vital Sign and ST Segment Trends
108
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
How VS Trends Work
Each active vital sign can be displayed graphically for time ranges of 30 minutes, and 1, 2, 4,
and 8 hours. The vital signs are HR, SpO
2
, SpCO, SpMet, CO
2
, Temp, and RR; and systolic,
diastolic, and mean pressures. Data is sampled every 30 seconds. If valid data is not available, a
blank space is substituted on the graph. NIBP values are plotted only when an NIBP
measurement is obtained. VS measurements are not averaged or filtered. No messages or
alarms occur based on changes in VS measurements.
Figure 29
EtCO2 Trend Graph
FIGURE LEGEND
1
First EtCO
2
measurement
2
Most recent EtCO
2
measurement
3
VS label