Summary of Screen Messages
246
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
MESSAGE
DESCRIPTION
CHECK PATIENT. IF NO PULSE,
PUSH ANALYZE
A potentially shockable rhythm has been detected when using
Advisory Monitoring.
CHECK PRINTER
The printer door is open, there is no paper in the printer, or
another printer malfunction exists.
CO2 AUTOZERO
EtCO
2
monitor is automatically performing a zero-point
calibration.
CO2 FILTERLINE BLOCKAGE
EtCO
2
FilterLine tubing is kinked or clogged; the message
appears after 30 seconds of unsuccessful purging.
CO2 FILTERLINE OFF
EtCO
2
FilterLine tubing is disconnected or is not securely
connected to the device.
CO2 FILTERLINE PURGING
EtCO
2
FilterLine tubing is kinked or clogged with liquid.
CO2 INITIALIZING
EtCO
2
monitor is performing a self-check.
CONNECT CABLE
Therapy cable is not connected when you press
CHARGE
,
PACER
, or
ANALYZE
.
CONNECT CHEST LEADS
A 12-lead ECG analysis was requested and precordial leads
are not connected to the patient.
CONNECT ECG LEADS
ECG electrodes or leads are disconnected.
CONNECT ELECTRODES
Therapy electrodes are disconnected.
CONNECTED TO
The device is connected via
Bluetooth
technology to another
Bluetooth
-enabled device. The name of the connected device
follows this message.
CONNECTING TO
The device is establishing communication with another
Bluetooth
-enabled device. The name of the target device
follows this message.
CPR: ADULT-AIRWAY X:Y
An option for CPR metronome. The patient is an adult for
whom an advanced airway has been established. The
specified C:V ratio will be used.
CPR: ADULT-NO AIRWAY X:Y
An option for CPR metronome. The patient is an adult for
whom an advanced airway has not been established. The
specified C:V ratio will be used.
CPR: YOUTH-AIRWAY X:Y
An option for CPR metronome. The patient is a youth (younger
than the age of puberty) for whom an advanced airway has
been established. The specified C:V ratio will be used.
CPR: YOUTH-NO AIRWAY X:Y
An option for CPR metronome. The patient is a youth (younger
than the age of puberty) for whom an advanced airway has
not been established. The specified C:V ratio will be used.
CURRENT FAULT
The comparison between delivered and selected pacing
current is out of tolerance.
DEMAND
Pacemaker is in Demand mode.
DEMO MODE
The device is in Demo mode and simulated patient data is
displayed.
DISARMING...
The energy charge is being removed internally.
ECG CABLE OFF
The device is printing and the ECG cable is removed.
ECG LEADS OFF
Multiple ECG electrodes are disconnected.
ENDING DEVICE SEARCH
The request for finding a
Bluetooth
device was stopped.
ENERGY DELIVERED
Energy transfer is complete.
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......