Monitoring Invasive Pressure
96
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Possible Patient Injury or Equipment Damage
Use only IP transducers that are specified for use with
this device. Protection of the device against defibrillator
discharge is dependent on the use of IP transducers
that are specified by Physio-Control.
Possible Lethal Arrhythmia
Ventricular fibrillation may be induced if the isoelectric
barrier of the transducer is disrupted. The isoelectric
barrier within the transducer may be disrupted if the
transducer body is damaged. Do not use a transducer
that is visibly damaged or leaking fluid.
Increased Intracranial Pressure
Do not use a continuous flush device with transducers
used for intracranial monitoring.
IP Monitoring
Two channels are available for invasive pressure monitoring, with default labels P1 and P2 and
the user-selectable labels shown in the following table.
Table 16
IP Labels and Descriptions
LABEL
DESCRIPTION
ART
Arterial Pressure
PA
Pulmonary Artery Pressure
CVP
Central Venous Pressure
ICP
Intracranial Pressure
LAP
Left Atrial Pressure
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......