Overview of the Shock Advisory System
The Shock Advisory System (SAS™) is an ECG analysis system built into the biphasic
LIFEPAK 15 monitor/defibrillator that advises the operator as to whether it detects a shockable
or nonshockable rhythm. This system makes it possible for individuals who are not trained to
interpret ECG rhythms to provide potentially lifesaving therapy to victims of ventricular fibrillation
or pulseless ventricular tachycardia.
The Shock Advisory System contains the following features:
•
Electrode Contact Determination
•
Automated Interpretation of the ECG
•
Operator Control of Shock Therapy
•
Continuous Patient Surveillance System (CPSS)
•
Motion Detection
The Shock Advisory System is active when the LIFEPAK 15 monitor/defibrillator is used as an
automated external defibrillator (AED). CPSS may be activated during monitoring.
Upon the user pressing the (shock) button, the LIFEPAK 15 monitor/defibrillator delivers the
shock therapy to the patient.
Electrode Contact Determination
The Shock Advisory System measures the patient's transthoracic impedance through the
therapy electrodes. If the baseline impedance is higher than a maximum limit, it determines that
the electrodes do not have sufficient contact with the patient or are not properly connected to
the AED. When this occurs, ECG analysis and shock delivery are inhibited. The AED advises the
operator to connect electrodes when there is insufficient electrode contact.
Automated Interpretation of the ECG
The Shock Advisory System recommends a shock if it detects the following:
•
Ventricular fibrillation—with a peak-to-peak amplitude of at least 0.08 mV.
•
Ventricular tachycardia—defined as having a heart rate of at least 120 beats per minute, QRS
width of at least 0.16 seconds, and no apparent P waves.
Pacemaker pulses may prevent advisement of an appropriate shock, regardless of the patient’s
underlying rhythm. The Shock Advisory System recommends no shock for all other ECG
rhythms including asystole, pulseless electrical activity, idioventricular rhythms, bradycardia,
supraventricular tachycardias, atrial fibrillation and flutter, heart block, premature ventricular
complexes, and normal sinus rhythms. These rhythms are specifically mentioned in the AHA
recommendations. The SAS does not continue analyzing the ECG after a
SHOCK ADVISED
decision is reached.
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......