Overview of the Shock Advisory System
256
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
present. If motion is detected, the ECG analysis is inhibited. The operator is advised by a
displayed message, a voice prompt, and an audible alert. After 10 seconds, if motion is still
present, the motion alert stops and the analysis always proceeds to completion. This limits the
delay in therapy in situations where it may not be possible to stop the motion. However, the
rescuer should remove the source of motion whenever possible to minimize the chance of
artifact in the ECG.
There are two reasons why ECG analysis is inhibited when the motion alert occurs, and why the
rescuer should remove the source of the motion whenever possible:
1.
Such motion may cause artifact in the ECG signal. This artifact can cause a nonshockable
ECG rhythm to look like a shockable rhythm. For example, chest compressions during
asystole can look like shockable ventricular tachycardia. Artifact can also cause a shockable
ECG rhythm to look like a nonshockable rhythm. For example, chest compressions during
ventricular fibrillation can look like an organized, and therefore nonshockable, rhythm.
2.
The motion may be caused by a rescuer’s interventions. To reduce the risk of inadvertently
shocking a rescuer, the motion alert prompts the rescuer to move away from the patient. This
will stop the motion and ECG analysis will proceed.
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......