Battery Maintenance
210
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Possible Equipment Damage
When storing the LIFEPAK 15 monitor/defibrillator for
an extended period of time, the battery should be
removed from the device.
Receiving New Batteries
New batteries do not arrive fully charged. Charge each new battery before use. Batteries may be
charged using any of the following devices:
•
Station Lithium-ion battery charger for use with the LIFEPAK 15 monitor/defibrillator
•
Mobile Lithium-ion battery charger for use with the LIFEPAK 15 monitor/defibrillator
•
REDI-CHARGE battery charger
•
AC power adapter for use with the LIFEPAK 15 monitor/defibrillator
•
DC power adapter for use with the LIFEPAK 15 monitor/defibrillator
Storing Batteries
Li-ion batteries self-discharge during storage.
If you store the battery:
•
Do not remove the Charge Before Use label to indicate that the battery has not yet been
charged.
•
Store batteries at temperatures between 20° to 25°C (68° to 77°F).
•
Charge the battery fully within one year of when you receive it. Fully recharge the battery
once per year thereafter.
Possible Loss of Power During Patient Care
Stored batteries lose charge. Failure to charge a stored
battery before use may cause device power failure
without warning. Always charge a stored battery before
placing it in active use.
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......