Specifications and Performance Characteristics
232
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Battery Operation Only:
Maximum Time from Charge to Shock Ready (Manual Mode):
Battery Status
Charge Time
Fully charged
200 J within 7 seconds, nominal
Fully charged, followed by 15
full-energy shocks
360 J within 10 seconds
Fully charged
360 J within 10 seconds
Maximum Time from Initiation of Analysis to Shock Ready (AED Mode):
Battery Status
Charge Time
Fully charged
200 J within 15 seconds, nominal
Fully charged, followed by 15
full-energy shocks
360 J within 30 seconds
Fully charged
360 J within 30 seconds
Maximum Time from Power-on to Shock Ready (Manual Mode):
Battery Status
Charge Time
Fully charged, followed by 15
full-energy shocks
360 J within 25 seconds
Maximum Time from Power-on to Shock Ready (AED Mode):
Battery Status
Charge Time
Fully charged, followed by 15
full-energy shocks
360 J within 40 seconds
Manual Mode
Energy Select
2, 3, 4, 5, 6, 7, 8, 9, 10, 15, 20, 30, 50, 70, 100, 125, 150, 175, 200,
225, 250, 275, 300, 325, and 360 joules
Synchronous
cardioversion
The maximum time delay between synchronization pulse and the
delivery of energy, once the output has been activated, is not more
than 60 msec. This time delay is measured from the peak of the QRS
to the peak of the defibrillator waveform.
Paddles Lead Off
Sensing
When using QUIK-COMBO electrodes, the device indicates Paddles
Leads Off if the resistive part of the patient impedance is greater
than 300 ±15%
Ω
, or if the magnitude of the patient impedance is
greater than 440 ±15%
Ω
.
Biphasic Waveform
Biphasic Truncated Exponential
The following specifications apply from 25 to 200
Ω
, unless otherwise
specified:
Energy Accuracy: ±1 joule or 10% of setting, whichever is greater,
into 50
Ω
;
±2 joules or 15% of setting, whichever is greater,
into 25-175
Ω
.
Voltage Compensation: Active when disposable therapy electrodes
are attached. Energy output within ±5% or ±1 joule, whichever is
greater, of 50
Ω
value, limited to the available energy which results in
the delivery of 360 joules into 50
Ω
.
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......