OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION
Target device is not functioning
•
Confirm that target device is on
and discoverable.
•
Confirm friendly name of target
device.
•
If message still appears, contact
the service provider for your
target device.
Bluetooth
module in
LIFEPAK 15 monitor may be
faulty
•
Contact qualified service
representative.
Unable to transmit data
for post-event review
using either direct
connection or
Bluetooth
connection
Post-event review software is not
installed on target device
•
Install CODE-STAT or
DT EXPRESS post-event review
software on target device.
Post-event review software is not
open and running on target
device
•
Make sure the target device is
running Device Communications
or the download wizard.
COM port is not configured for
incoming data on target device
•
Configure COM port on target
device.
LIFEPAK 15 monitor not selected
in download wizard on target
device
•
Open download wizard on target
device and select the
LIFEPAK 15 monitor.
BLUETOOTH
UNAVAILABLE
message
appears
Bluetooth
module in
LIFEPAK 15 monitor not
responding
•
Turn LIFEPAK 15 monitor off and
back on.
•
If message still appears,
Bluetooth
module may be faulty.
Contact qualified service
representative.
BLUETOOTH DEVICE
NOT FOUND
message
appears
Unable to locate
Bluetooth
device
•
Verify target device is ready to
receive transmissions.
•
Set
SEARCH FILTER
to
OFF
and
then select
FIND DEVICES
again.
UNKNOWN DEVICE
message appears
Bluetooth
name discovery failed
or timed out before the device
name was obtained
•
Verify name of target device.
•
Verify target device is ready to
receive transmissions.
•
Attempt to retransmit.
Unable to transmit using
a gateway device that
has a functioning direct
connection or
Bluetooth
connection
Transmission sites are not set up
in LIFEPAK 15 monitor
•
Define transmission sites. Each
site name must exactly match the
name of the target device. See
LIFEPAK 15 Monitor/Defibrillator
Setup Options
provided with your
device.
Transmission site names in
LIFENET System do not match
site names in
LIFEPAK 15 monitor
•
Check site names in LIFENET
System.
Cellular communication is not
working between the gateway
and transmission sites
•
Use alternate method to
communicate patient data.
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......