Specifications and Performance Characteristics
240
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Table 43
Setup Options Factory Default Settings
MENU
MENU/ITEM
FACTORY DEFAULT
SETTINGS
General
Language
(Country Specific)
Code Summary
Long
Trend Summary
Off
Site Number
000
Device ID
"LP15" + last 4 digits of serial number, for example,
LP151234
Auto Log
On
Line Filter
60 Hz
Timeout Speed
30 seconds
Manual mode
Sync After Shock
Off
Pads Default
200 (joules)
Energy Protocol
Inactive
Internal Default
10 (joules)
Voice Prompts
On
Shock Tone
On
Manual Access
Manual / Direct
Set Passcode
0000
AED mode
Energy Protocol
200–300–360
Auto Analyze
Off
Motion Detection
On
Pulse Check
Never
CPR
CPR Time 1
120 seconds
CPR Time 2
120 seconds
Initial CPR
Off
Initial CPR Time
120 seconds
Preshock CPR
Off
CPR
Metronome
Metronome
On
Adult - No Airway
30:2
Adult - Airway
10:1
Youth - No Airway
15:2
Youth - Airway
10:1
Pacing
Rate
60 PPM
Current
0 mA
Mode
Demand
Internal Pacer
Detection Off
Monitoring
Channels
Default Set
Set 1
Set 1
Channel 1
ECG Lead II
Channel 2
None
Channel 3
None
Continuous Data
ECG Channel 1
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......