Chapter 6
| Paddle Accessory Options
©2007-2015 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
153
To remove QUIK-COMBO therapy electrodes from the patient:
1.
Slowly peel back the therapy electrode from the edge, supporting the skin as shown.
Figure 37
Removing Therapy Electrodes from Skin
2.
Clean and dry the patient’s skin.
3.
When applying new electrodes, adjust the positions slightly to help prevent skin burns.
4.
Close the protective cover on the therapy cable connector when the cable is not in use.
Cleaning and Inspection
QUIK-COMBO electrodes are not sterile or sterilizable. They are disposable and are for a single
patient application. Do not autoclave, gas sterilize, immerse in fluids, or clean electrodes with
alcohol or solvents.
Include daily inspection of the QUIK-COMBO therapy electrode package as part of your
defibrillator test routine. Daily inspection helps ensure that the therapy electrode has not
exceeded the electrode package Use By date and is ready for use when needed. For more
information about daily inspection and testing, see the Operator’s Checklist in the back of this
manual.
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......