Chapter 7
| Data Management
©2007-2015 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
167
Memory Capacity
The LIFEPAK 15 monitor/defibrillator retains data for two or more patients when you switch
power off or remove the batteries. The number of patient reports that the
LIFEPAK 15 monitor/defibrillator can store depends on various factors, including the number of
displayed waveforms, the duration of each use, and the type of therapy. The total capacity
is 360 minutes of continuous ECG, 90 minutes of continuous data from all channels,
or 400 single waveform events. The maximum memory capacity for a single patient includes up
to 200 single waveform reports and 90 minutes of continuous ECG. When the defibrillator
reaches the limits of its memory capacity, the defibrillator deletes an entire Patient Record using
a "first in, first out" priority to accommodate a new Patient Record. Deleted Patient Records
cannot be retrieved.
Managing Current Patient Records
You can add specific patient information to a current Patient Record. For more information, see
Entering Patient Data (on page 42).
How to Print a Current Report
To print a current report:
1 Press
OPTIONS
. The Options menu appears.
2 Select
. The Options/Print menu
appears.
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......