Monitoring the ECG
48
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Selecting ECG Lead
The LIFEPAK 15 monitor/defibrillator includes two methods for selecting or changing the ECG
lead.
To select or change the displayed ECG lead using the
LEAD
button:
1.
Press
LEAD
. If any ECG lead currently
appears on the Home Screen, the lead
changes to
PADDLES
. If
PADDLES
lead is
currently displayed, the lead changes to
Lead
II
.
2.
While the
LEAD
menu is displayed, press
LEAD
again or rotate the
SPEED DIAL
to
the desired lead.
Note
: If lead sets are predefined for
Channels 2 and 3, the lead sets show on the
menu. The ECG cable that is connected to
the device, such as 3-lead or 5-wire,
determines the leads you can select. For
information about defining lead sets, see the
LIFEPAK 15 Monitor/Defibrillator Setup
Options
provided with your device.
To select or change the displayed ECG lead using the
SPEED DIAL
:
1.
For the primary ECG, outline and select
CHANNEL 1
and then select
LEAD
.
2.
Rotate the
SPEED DIAL
to the desired
ECG lead.
3.
Press the
SPEED DIAL
to select the ECG
lead.
4.
Repeat this procedure to select or
change displayed ECG waveforms for
Channels 2 and 3.
Note:
The ECG shows dashed lines until the electrodes are connected to the patient.
Note:
When the
VF/VT ALARM
is ON, you are limited to
PADDLES
lead or Lead
II
in Channel 1.
See Setting Alarms (on page 39).
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......