Troubleshooting Tips
186
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Troubleshooting Tips
Table 32
Troubleshooting Tips for Data Transmission
OBSERVATION
POSSIBLE CAUSE
CORRECTIVE ACTION
Bluetooth
icon on
LIFEPAK 15 monitor has
red X across it
WIRELESS
is set to
OFF
in the
Bluetooth
Setup menu
•
Set
WIRELESS
to
ON
.
•
If red X remains,
Bluetooth
module in LIFEPAK 15 monitor
may be faulty. Contact qualified
service representative.
WIRELESS
is set to
OFF
in the
setup options, so the
WIRELESS
default is
OFF
each time the
LIFEPAK 15 monitor is turned on
•
Change
WIRELESS
setup option.
See
LIFEPAK 15 Monitor/Defibrillator
Setup Options
provided with your
device.
•
If red X remains,
Bluetooth
module in LIFEPAK 15 monitor
may be faulty. Contact qualified
service representative.
Bluetooth
module in
LIFEPAK 15 monitor may be
faulty
•
Contact qualified service
representative.
Bluetooth
LED is not
illuminated
Target device is off or cannot
communicate with the
LIFEPAK 15 monitor
•
Confirm that target device is on
and discoverable.
•
See the operating instructions for
your target device.
Bluetooth
module in
LIFEPAK 15 monitor may be
faulty
•
If other troubleshooting is
unsuccessful, contact qualified
service representative.
LIFEPAK 15 monitor
does not automatically
connect to last
connected device
Target device is off or cannot
communicate with the
LIFEPAK 15 monitor
•
Confirm that target device is on
and discoverable.
Last connection to target device
may have occurred when the
LIFEPAK 15 monitor was in a
different mode
•
Confirm that
OUTPUT PORT
is
set to
BLUETOOTH WIRELESS
.
•
Select
FIND DEVICES
and
establish a new connection.
Device does not connect
to last connected device
after
WIRELESS
is set to
ON
Bluetooth
menu is displayed,
which prevents discovery of
devices
•
Press
HOME SCREEN
to exit
menu and allow
LIFEPAK 15 monitor to find last
connected device.
UNABLE TO CONNECT
message appears
LIFEPAK 15 monitor cannot
establish wireless connection.
Target device may not have the
necessary software application or
cannot accept data.
•
Verify target device is ready to
receive transmissions.
•
Attempt to retransmit.
Unable to find a
particular
Bluetooth
device, or
BLUETOOTH
DEVICE NOT FOUND
message appears
Search filter may be on and target
device does not have a PSC
prefix
•
Confirm that target device is on
and discoverable.
•
Confirm friendly name of target
device.
•
Set
SEARCH FILTER
to
OFF
and
then select
FIND DEVICES
again.
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......