Chapter 5
| Therapy
©2007-2015 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
123
Note:
If you do not press the (shock)
button within 60 seconds, or the
SPEED DIAL
is pressed to cancel charging, the
defibrillator disarms and the
DISARMING
message appears.
After a shock is delivered, the
START CPR
prompts occur. A countdown timer (min:sec
format) continues for the duration specified in
the
CPR TIME 1
setup option.
Note:
The CPR metronome automatically
provides audible compression “tocks” and
ventilation prompts or tones only during CPR
intervals at a ratio of 30:2. To silence the
metronome, press
CPR
. To un-silence the
metronome, press
CPR
again.
When the CPR countdown time ends, the
PUSH ANALYZE
prompts occur. These
prompts repeat every 20 seconds until you
press
ANALYZE
.
No Shock Advised
The following prompts occur if no shock is advised:
If the AED detects a nonshockable
rhythm, the
NO SHOCK ADVISED
prompts
occur. The defibrillator does not charge,
and no shock can be delivered.
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......