Monitoring the ECG
50
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
Adjusting the Systole Volume
To adjust the systole beep volume, use the
SPEED DIAL
to outline and select the
HR
area on the
Home Screen.
The following menu appears:
1.
Press the
SPEED DIAL
to select
QRS
VOLUME
.
2.
Rotate the
SPEED DIAL
to the desired
volume.
3.
Press the
SPEED DIAL
to set the volume.
Note:
The volume is reset to OFF each time the device is turned off.
Monitoring Using Paddle Accessories
To monitor ECG using paddles, you can use either QUIK-COMBO therapy electrodes or
standard (hard) paddles. For more information about paddle accessories, see Paddle Accessory
Options (on page 149).
Anterior-Lateral Placement
Anterior-lateral placement is the only placement that should be used for ECG monitoring
using paddle accessories.
To place the therapy electrodes or paddles:
1.
Place either the
therapy electrode or
APEX
paddle lateral to the patient's left nipple in
the midaxillary line, with the center of the electrode in the midaxillary line, if possible, as
shown in the following figure.
QUIK-COMBO Therapy Electrodes
Standard Paddles
Figure 15
Anterior-Lateral Placement
2.
Place the other therapy electrode or
STERNUM
paddle on the patient’s upper right torso,
lateral to the sternum and below the clavicle, as shown in the preceding figure.
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......