Appendix B
| Screen Messages
©2007-2015 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
247
MESSAGE
DESCRIPTION
ENERGY FAULT
The comparison between stored and selected energy is out of
tolerance.
ENTER MANUAL MODE?
One of the Manual mode access buttons was pressed and the
confirmation screen is set up to appear.
EXCESSIVE NOISE - 12-LEAD
CANCELLED
Noise is detected for longer than 30 seconds that is too great
to record a 12-lead ECG report.
IF NO PULSE, PUSH ANALYZE
Follows a CPR interval, if a
PULSE CHECK
setup option other
than
NEVER
is selected.
IF NO PULSE, START CPR
Follows delivery of a shock or
NO SHOCK ADVISED
prompt, if
a
PULSE CHECK
setup option other than
NEVER
is selected.
IF YOU WITNESSED THE ARREST,
PUSH ANALYZE
Initial CPR message that follows
START CPR
prompt, to
remind user to deliver a shock immediately if the user
witnessed the arrest.
LA LEADS OFF
ECG electrode "LA" is disconnected.
LAST CONNECTED TO
When
Bluetooth
connectivity is installed and this device
previously connected to a target device, the name of the
target device appears after this message.
LL LEADS OFF
ECG electrode "LL" is disconnected.
LOST BLUETOOTH CONNECTION Communication with
Bluetooth
device has been interrupted.
LOST DIRECT CONNECTION
Communication via direct connection has been interrupted.
MAINTENANCE DUE
Reminder message that appears at the interval that is set in
Service mode. Message continues to appear until reset or
turned off.
MANUAL MODE DISABLED
Access to Manual mode from AED mode has been restricted.
MOTION DETECTED!/STOP
MOTION!
Motion was detected during ECG analysis.
NIBP AIR LEAK
NIBP cuff applied too loosely or there is a leak in cuff/monitor
pneumatic system.
NIBP CHECK CUFF
NIBP cuff is not connected to patient or device.
NIBP FAILED
NIBP monitor cannot establish zero-pressure reference.
NIBP FLOW ERROR
NIBP pneumatic system is not maintaining stable cuff
pressure.
NIBP INITIALIZING
NIBP requested while NIBP module is still initializing.
NIBP MOTION
Patient extremity moved too much for the NIBP monitor to
accurately complete the measurement.
NIBP OVERPRESSURE
NIBP cuff pressure exceeded 290 mmHg.
NIBP TIME OUT
NIBP monitor did not complete a measurement
in 120 seconds.
NIBP WEAK PULSE
The monitor did not detect any pulses.
NO SHOCK ADVISED
The defibrillator did not detect a shockable rhythm.
NO SITES DEFINED
Device is attempting to transmit using
Bluetooth
connection,
but no associated destinations have been defined.
NOISY DATA! PRESS 12-LEAD TO
ACCEPT
Monitor detects excessive signal interference while acquiring
data. Press
12-LEAD
to override the message and acquire 12-
lead ECG with noise.
NON-DEMAND
Pacemaker is in Nondemand (asynchronous) mode.
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......