To perform automated external defibrillation:
1.
Verify that the patient is in cardiopulmonary arrest (unconscious, pulseless, not breathing
normally).
2.
Press
ON
.
3.
Prepare the patient for electrode placement (see Patient Skin Preparation (on page 117)).
The
CONNECT ELECTRODES
prompts
occur until the patient is connected to the
AED. If possible, place the patient on a hard
surface away from standing water.
4.
Connect the therapy electrodes to the therapy cable and confirm cable connection to the
defibrillator.
5.
Apply the therapy electrodes to the patient’s chest in the anterior-lateral position (see
Anterior-Lateral Placement (on page 117)).
The
PUSH ANALYZE
prompts occur when
the patient is properly connected to the
AED.
6.
Press
ANALYZE
to initiate the analysis. Stop CPR.
Possible Misinterpretation of Data
Do not move the AED during analysis. Moving the AED
during analysis may affect the ECG signal resulting in
an inappropriate
SHOCK
or
NO SHOCK ADVISED
decision. Do not touch the patient or the AED during
analysis.
Summary of Contents for LIFEPAK 15
Page 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 2: ......
Page 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Page 8: ......
Page 10: ......
Page 14: ......
Page 20: ......
Page 44: ......
Page 46: ......
Page 112: ......
Page 114: ......
Page 148: ......
Page 150: ......
Page 160: ......
Page 174: ......
Page 190: ......
Page 198: ......
Page 200: ......
Page 222: ......
Page 244: ......
Page 250: ......
Page 252: ......
Page 258: ......
Page 262: ......
Page 268: ......