Top Case Assemblies
4
-
38
Removal and Replacement
SpO
2
PCA
The SpO
2
PCA is located next to the ECG and SpO
2
connectors.
Removal
1. Disconnect and remove the ECG connector. See "ECG Connector" on
page 4-40.
2. Disconnect and remove the SpO
2
connector. See "SpO
page 4-42.
3. Note the orientation and placement of the flex circuit from the SpO
2
PCA
to the Interface PCA. Disconnect this flex circuit at both ends. These con-
nectors have a sliding bar latch - slide out toward the edge of the PCA to
release, then pull out the flex.
4. Remove the SpO
2
PCA and discard the flex circuit to the Interface PCA.
Figure 4-22 Removing the Screws - SpO
2
PCA
ECG Connector
SpO
2
Connector
Interface PCA
Interface PCA Connectors:
SpO
2
ECG
Flex Circuit - SpO
2
PCA to Interface
PCA
Ferrite
SpO
2
PCA