M3500B/M5500B Service Manual
4
-
55
Top Case Assemblies
4
Speaker
These instructions describe how to remove and replace the speaker assembly.
Removal
1. Disconnect the speaker from the Control PCA. Refer to Figure 4-18 to
identify the speaker connector.
2. Remove the 2 screws and clamps from the speaker. See Figure 4-30.
3. Peel up the tape holding the speaker wires in position. Lift the speaker
and its wires out of the case.
Figure 4-30 Removing the Speaker
Replacement
1. Place the new speaker into position in the case.
2. Install both clamps and both screws.
3. Lay the wires in the same position in the case as were the original wires,
and secure with the original tape (or with other plastic electrical tape).
4. Connect the speaker to the Control PCA.
Display Window
Tape
Speaker
Wires
Speaker
Clamp