Test and Inspection Matrix
2
-
4
Performance Verification and Safety Tests
User Interface
Tests (U)
•
In Diagnostic Mode, run
the following tests
(page 2-19):
If all responses as expected:
Example: U:p
•
Controls Test
•
All keys respond as
expected
•
Display Test
•
Display goes dark, then
light. Black vertical bar
scrolls across the screen
from left to right.
•
Audio Test
•
Alerts, alarms, and tones
are clearly heard
•
Printer Test
•
Print quality is adequate;
no stray marks or lines
•
Print speed: 25 mm + 5%
(1.25mm)
then User Interface test
passes
ECG Tests (E)
In Diagnostic Mode, run the
ECG Tests (page 2-22):
If all data within limits, all
checks pass:
Example:E:p
•
Status messages (lead,
pad, DSP)
•
"Good" displayed for all
three status messages
PCI measurement
If all PCI measurements are
within the following ranges:
•
PCI - Test Load attached
•
50 ± 30
Ω
•
PCI - Pads open
•
> 2000
Ω
then ECG test passes
Pacing Test (P)
In Diagnostic Mode, run the
Pacing Test (page 2-25):
If measurement is within the
following limit:
Example: P:p
•
200 mA
•
200 mA± 20 mA
then Pacing test passes
Table 2-1 Performance Verification and Safety Tests (Continued)
Test Group Name
Test or Inspection to
Perform
Expected Test Results
Data to Record
P (pass) or F (fail)
Summary of Contents for Heartstart XLT
Page 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Page 2: ......
Page 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Page 6: ......
Page 16: ......
Page 20: ......
Page 52: ......
Page 216: ......
Page 223: ......
Page 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...