M3500B/M5500B Service Manual
2
-
3
Test and Inspection Matrix
2
Test and Inspection Matrix
The matrix in Table 2-1
summarizes performance verification tests and
inspections for the M3500B/M5500B; including test name, test or inspection
to perform, expected test results, and data to record.
Table 2-1 Performance Verification and Safety Tests
Test Group Name
Test or Inspection to
Perform
Expected Test Results
Data to Record
P (pass) or F (fail)
Visual Inspection
(VI)
Inspect unit, accessories,
cables, etc. as described on
page 2-12.
•
If no unusual damage, no
corrosion, no missing
items,
then Visual Inspection
passes
Example VI:p
Functional
Checks (F)
In normal Operating Mode,
perform the following func-
tional checks:
•
If all functions respond as
expected:
Example F:p
•
ECG (page 2-13).
•
Waveform clear on dis-
play; HR correct on display;
HR alarm works. Leads off,
pads off indicators as
expected.
•
Shock Advisory (page 2-
•
Shock Advised only when
appropriate
•
Synchronized Cardiover-
sion (page 2-14).
•
Shock delivered with cor-
rect timing (< 60msec)
•
SpO
2
(page 2-15).
•
95% -100%
then Functional test passes
Extended Self
Test (X)
In Diagnostic Mode, run the
Extended Self Test (page 2-
17). Includes Data Card Test
and time/date check.
If "Pass" reported on all
tests applicable to the device
configuration and options,
then Extended Self test
passes
Example X:p
Summary of Contents for Heartstart XLT
Page 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Page 2: ......
Page 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Page 6: ......
Page 16: ......
Page 20: ......
Page 52: ......
Page 216: ......
Page 223: ......
Page 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...